Sysmex XN-L Series
Description
As the global leaders in hematology, we consider it our responsibility to offer you solutions that make your work easier, more efficient and more effective. In 2011 we introduced the benchmark in hematology – our flagship XN-Series. For some, such a comprehensive solution may be too large. Now, with the introduction of the XN-L Series, XN quality and Sysmex’s knowledge and support are within your grasp.
Whether you want to upgrade from a 3-part differential device or replace an existing analyzer, all XN-L models are full 5-part differential analyzers that offer the quality
of larger devices at an excellent price: quality ratio.
The new XN-L Series comes in three models: the XN-350, the XN-450 and the XN-550. Each is also a perfect secondary analyzer, giving you confidence even at times when your primary analyzer is unavailable
Sysmex XN 350
Dimensions/weights W x H x D [mm/kg] |
XN-350: 450 x 510 x 460 / approx. 35 kg |
Throughput | up to 60 samples/h in WB mode up to 70 samples/h in WB mode with the optional Speed-up license |
Aspiration volume | 25 µL in WB mode, 70 µL in PD and BF mode |
Analysis parameters | Whole Blood WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, RDW-SD, RDW-CV, PDW, MPV, P-LCR, PCT, NEUT#, LYMPH#, MONO#, EO#, BASO#, NEUT%, LYMPH%, MONO%, EO%, BASO%, IG#, IG% With RET (optional) Body Fluids with XN-BF (optional) |
Measurement principles | WBC DIFF/RET: Fluorescence Flow Cytometry WBC: Flow Cytometry RBC/PLT: DC Impedance method with hydrodynamic focussing HGB: Cyanide-free SLS method |
Data storage capacities | Results: 100,000 samples Patient information: 10,000 records Calibration history: 20 times/ module QC files: 99 files including XbarM |
User Maintenance | Daily shutdown Weekly cleaning sequence with Cellclean or Cellclean Auto |
Sysmex XN-450
Dimensions/weights W x H x D [mm/kg] |
XN-450: 450 x 440 x 460 / approx. 35 kg |
Throughput | up to 60 samples/h in WB mode up to 70 samples/h in WB mode with the optional Speed-up license |
Aspiration volume | 25 µL in WB mode, 70 µL in PD and BF mode |
Analysis parameters | Whole Blood WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, RDW-SD, RDW-CV, PDW, MPV, P-LCR, PCT, NEUT#, LYMPH#, MONO#, EO#, BASO#, NEUT%, LYMPH%, MONO%, EO%, BASO%, IG#, IG% With RET (optional) Body Fluids with XN-BF (optional) |
Measurement principles | WBC DIFF/RET: Fluorescence Flow Cytometry WBC: Flow Cytometry RBC/PLT: DC Impedance method with hydrodynamic focussing HGB: Cyanide-free SLS method |
Data storage capacities | Results: 100,000 samples Patient information: 10,000 records Calibration history: 20 times/ module QC files: 99 files including XbarM |
User Maintenance | Daily shutdown Weekly cleaning sequence with Cellclean or Cellclean Auto |
Sysmex XN-450
Dimensions / weights W x H x D [mm/kg] |
XN-550: 450 x 450 x 660 / approx. 53 kg |
Throughput | up to 60 samples/h in WB mode up to 70 samples/h in WB mode with the optional Speed-up license |
Aspiration volume | 25 µL in WB mode, 70 µL in PD and BF mode |
Analysis parameters | Whole Blood WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, RDW-SD, RDW-CV, PDW, MPV, P-LCR, PCT, NEUT#, LYMPH#, MONO#, EO#, BASO#, NEUT%, LYMPH%, MONO%, EO%, BASO%, IG#, IG% With RET (optional) Body Fluids with XN-BF (optional) |
Measurement principles | WBC DIFF/RET: Fluorescence Flow Cytometry WBC: Flow Cytometry RBC/PLT: DC Impedance method with hydrodynamic focussing HGB: Cyanide-free SLS method |
Data storage capacities | Results: 100,000 samples Patient information: 10,000 records Calibration history: 20 times/ module QC files: 99 files including XbarM |
User Maintenance | Daily shutdown Weekly cleaning sequence with Cellclean or Cellclean Auto |