DYNEX DSX

ipsum dolor sit amet, consectetur adipiscing elit, sed do eiusmod tempor incididunt ut labore et dolore magna aliqua. Egestas purus viverra accumsan in nisl nisi. Arcu cursus vitae congue mauris rhoncus aenean vel elit scelerisque. In egestas erat imperdiet sed euismod nisi porta lorem mollis.

The DYNEX DSX® is the next stage of our fully automated 4-plate ELISA processing system. This breakthrough technology has transformed labs and medical facilities across the country with its state of the art design, seamless automation and intuitive interface.

DSX® is a tried-and-tested automated open system capable of performing several assays per plate simultaneously, providing optimized efficiency and speed. The system uses a perfectly synchronized system to eliminate plate drift and ensure premium consistency across four different plate incubators. As a result, you can count on reliable quality while reaching peak performance speeds that once seemed unattainable.

The bar has been raised. Busy clinical and medium-throughput private and industrial labs require advanced automation that can handle virtually every ELISA immunoassay, all in a reliable, cost-effective package. Look no further than the DSX®.

4 independent plate incubators help manage testing workload with reduce labor, walk-away capability to maximize lab personnel efficiency.

The Worklist Load Wizard walks you through the process of setting up the DSX® worklist, graphically showing where to place reagents, samples, and places at the beginning of each run.

The DSX® delivers everything needed to ensure the rigorous, repeatable analyses required in critical applications including immunology, infectious disease, hormone, allergy, blood-chemistry, drugs-of-abuse testing, food safety, forensics and environmental testing.

For the past two years DYNEX DSX® has achieved 98.2% uptime when mean time between failure target is set at 250 days.

With over 3,500 DYNEX DSX® systems installed worldwide, DYNEX DSX® has been cited recently in 38 scientific journal publications for studies in blood borne pathogens, Quantiferon Gold Plus, autoimmune, sexually transmitted diseases, tick-borne diseases and others.

 

Dimensions Metric

Width: 106 cm

Depth: 91 cm

Height: 80 cm

Footprint: 0.98 sq m

Bench Weight: 136 kg (max.)

Ship Weight: 244 kg (max.)

Physical Specifications

Voltage: 100 – 240 V automatic conversion

Frequency: 50/60 Hz

Power Consumption: < 800 VA (online UPS recommended)

Power Supply Requirements

Photometric Range: 0.000 – 3.000 OD

Spectral Range: 405 nm – 690 nm

Filter Slots: 6 wavelength

Precision: ± 0.010 OD at 0.000 – 0.500 OD

< 1% CV at 0.501 – 2.000 OD

< 1.5% at 2.001 – 2.500 OD

Accuracy: ± 0.01 OD or 2.5%

(0.000 – 3.000 OD) whichever is greater

Read Time: < 10 seconds, single wavelength

< 20 seconds, dual wavelength

Reader Specifications

Manifold Configuration: 8-way

Programmable Volumes: 50 µL – 999 µL

Wash Buffer Capacity: 4 wash bottles at 2.0 L, with level-sensing

Residual Wash Volume: < 3 ?L per well with a dual-axis sweep in a

at-bottom plate

Dispense Precision: ? 5% CV (with 300 ?L in a 96-well plate)

Waste Container: 8 L with waste full sensor

Washer Specifications

Number of Plates: 4

Sample Capacity: 96 (98 positions available)

Continuous Sample Loading: Yes

Sample Tube Size: 10 mm – 16 mm diameter

40 mm – 100 mm height

Number of Reagents: Up to 24 (25 mL tubes)

Number of

Standard/Control Bottles: 33 (1.5 mL vials)

Assays per Plate: 1 assay per strip or up to 12 assays per plate

Dilution Wells Capacity: Up to 192 deep-well strips (24 x 8)

Sample Tip Capacity: 432 (4 x 108)

Reagent Tip Capacity: 41 tips

System Specifications

Sample Tip Size: 300 ?L

Sample Pipetting Volume: 10 ?L – 250 ?L single-shot

25 ?L – 100 ?L multi-shot

Sample Pipetting Precision: ? 3% CV at any operating volume above

10 ?L (single-shot mode)

Sample Pipetting Accuracy: ± 2% of target volume at any operating

volume above 10 ?L (single-shot mode)

Estimated Cycle Time for Sample

Pickup to Delivery on Plate: < 8 seconds^

Time to Dispense: 19 minutes (typical)^

50 ?L of 96 samples to plate from sample

tubes or deep-well plates

Sampling Time w/Dilutions: < 26 minutes (typical)^

Dilution Range: 1 part in 190 one-stage dilution

1 part in 36,100 two-stage dilution

Sample Pipetting Specications

Liquid Level Sensing: Yes (reagents, controls and samples)

Level Sensor System: Pressure dierential

Clot Detection: Yes

Dispense Anomaly Detection: Yes

Tip Detection: Yes

Electronic Signature Pipetting: Yes

Alarms: Yes

Process Security

Controlling Software: REVELATION DSX®

Work Protocols (Assays): Unlimited

Data Processing: Quantitative and qualitative

Levey-Jennings: Yes

Westgard Rules: Yes

Process Reporting: Event log + error log

Automatic Error Recovery: Yes

Password Access Control: Yes

Software

Reagent Tip Size: 1,300 µL

Reagent Pipetting Volume: 25 ?L – 1,000 µL

Reagent Pipetting Precision: ? 3% CV at 10 shots at any volume in

the operating range above 50 µL

Reagent Pipetting Accuracy: ± 2% of target volume at 50 µL or greater

in operating range (single-shot mode)

Reagent Pipetting Specications

Number of Incubators: Up to 4

Temperature Range: RT + 7° C to 50° C (elevated incubator)

RT + 5° C (ambient incubator)

Temperature Accuracy: ± 1° C

Shaking: > 15 Hz periodic or continuous

Temperature Monitoring: Yes